CISA

Flow Neuroscience FL-100

From Cybersecurity and Infrastructure Security Agency ↗

Successful exploitation of this vulnerability could allow an attacker within Bluetooth range to manipulate brain stimulation parameters and override safety limits.

The following versions of Flow Neuroscience FL-100 are affected:

Flow Neuroscience FL-100

Halo Neuroscience FL-100

Vendor

Equipment

Flow Neuroscience

Flow Neuroscience FL-100

Use of Hard-coded Credentials

Critical Infrastructure Sectors: Healthcare and Public Health

Countries/Areas Deployed: Worldwide

Company Headquarters Location: Sweden

An undocumented hard-coded credential, shared by all device units, is authorized to bypass authentication. This allows an attacker within Bluetooth range to arbitrarilymanipulate brain stimulation parameters and state.

Flow Neuroscience FL-100