CISA

Pulsetto Vagus Nerve Stimulator

From Cybersecurity and Infrastructure Security Agency ↗

Successful exploitation of this vulnerability could allow an attacker to use hidden commands to disable electrical safety mechanisms or modify other stimulation output settings.

The following versions of Pulsetto Vagus Nerve Stimulator are affected:

Pulsetto Vagus Nerve Stimulator vers:all/* (CVE-2026-18844)

Vendor

Equipment

Pulsetto

Pulsetto Vagus Nerve Stimulator

Hidden Functionality

Critical Infrastructure Sectors: Healthcare and Public Health

Countries/Areas Deployed: Worldwide

Company Headquarters Location: Lithuania

The firmware of the affected product accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.

Pulsetto Vagus Nerve Stimulator

MitigationPulsetto has...