CISA

Apollo Pharmacy Blood Glucose Monitoring System APG-01 BT

From Cybersecurity and Infrastructure Security Agency ↗

Successful exploitation of these vulnerabilities could allow an attacker to obtain sensitive health-related information and prevent legitimate users from establishing a connection with the device.

The following versions of Apollo Pharmacy Blood Glucose Monitoring System APG-01 BT are affected:

Blood Glucose Monitoring System (Model No. APG-01 BT) 0x0110_v1.1.0 (CVE-2026-50034, CVE-2026-52866)

Vendor

Equipment

Apollo Pharmacy

Apollo Pharmacy Blood Glucose Monitoring System APG-01 BT

Cleartext Transmission of Sensitive Information, Missing Authorization

Critical Infrastructure Sectors: Healthcare and Public Health

Countries/Areas Deployed: India

Company Headquarters Location: India

An attacker within BLE communication range can passively intercept wireless traffic and obtain sensitive health-related information, including glucose measurement values.

Apollo Pharmacy Blood Glu...